Why is risk management the most cited QMSR compliance violation?
Almost every company has a risk management file, but almost nobody has a risk management process. The risk file gets built during design, approved, and then placed in a drawer. Companies collect complaints and open non-conformances over five years, but none of this information flows back into the risk file. Under ISO 13485, risk management runs through product realization as its own documented process, so citations are not that your risk file is wrong, but that you don't have a risk management process.
Full answer and detailsRead the whole answer, key facts and every other question from this episode
In short
Almost every company has a risk management file, but almost nobody has a risk management process.
Nikhil Mangali, Vice President of Quality at Capstone Medical · #474: Behind the QMSR Audits: The Top 5 FDA Citations & Risk Management Pitfalls · Global Medical Device Podcast powered by Greenlight Guru · answer at 3:00
Key takeaways
- Almost every company has a risk management file, but almost nobody has a risk management process.
Quote to remember
“Your own risk documentation tells you that. Your own file is now the navigation instrument for an investigator.”
— Nikhil Mangali
The full answer · about 5 min
Almost every company has a risk management file, but almost nobody has a risk management process. The risk file gets built during design, approved, and then placed in a drawer. Companies collect complaints and open non-conformances over five years, but none of this information flows back into the risk file. Under ISO 13485, risk management runs through product realization as its own documented process, so citations are not that your risk file is wrong, but that you don't have a risk management process.
Key facts
- QMSR effective for seven months at time of episode recording
- Proposed rule released three and a half years before QMSR implementation
- Risk management process must run through product realization under ISO 13485 Clause 7.3, not just at design validation
- FDA compliance program organizes inspection around six QMS areas plus four FDA-specific requirements
- July 2024: Nikhil Mangali promoted to Vice President of Quality at Capstone Medical
People, companies and topics mentioned
More answers from this episode
- 0:54What are the top 5 FDA citations from QMSR audits according to real inspection data?
- 5:15What specific supplier control issues are FDA citing under QMSR?
- 14:20How should legacy medical devices with older DHF files approach QMSR compliance?
- 17:45What are the three layers of DHF to ISO 13485 mapping and what do most companies miss?
- 20:10What is the four-phase gap assessment process for QMSR compliance?